Microneedle systems have been widely used in health monitoring, painless drug delivery, and medical cosmetology. Although many studies on microneedle materials, structures, and applications have been conducted, the applications of microneedles often suffered from issues of inconsistent penetration rates due to the complication of skin-microneedle interface. In this study, we demonstrated a methodology of determination of transdermal rate of metallic microneedle array through impedance measurements-based numerical check screening algorithm. Metallic sheet microneedle array sensors with different sizes were fabricated to evaluate different transdermal rates. In vitro sensing of hydrogen peroxide confirmed the effect of transdermal rate on the sensing outcomes. A FEM simulation model of a microneedle array revealed the monotonous relation between the transdermal state and test current. Accordingly, two methods were first derived to calculate the transdermal rate from the test current. Firs...